Loading...
Thumbnail Image
Item

Assessment of serum total 25-hydroxyvitamin D assay commutability of Standard Reference Materials and College of American Pathologists Accuracy-Based Vitamin D (ABVD) Scheme and Vitamin D External Quality Assessment Scheme (DEQAS) materials: Vitamin D Standardization Program (VDSP) Commutability Study 2

Camara, Johanna E.
Wise, Stephen A.
Hoofnagle, Andrew N.
Williams, Emma L.
Carter, Graham D.
Jones, Julia
Burdette, Carolyn Q.
Hahm, Grace
Nalin, Federica
Kuszak, Adam J.
... show 10 more
Other Titles
Abstract
An interlaboratory study was conducted through the Vitamin D Standardization Program (VDSP) to assess commutability of Standard Reference Materials® (SRMs) and proficiency testing/external quality assessment (PT/EQA) samples for determination of serum total 25-hydroxyvitamin D [25(OH)D] using ligand binding assays and liquid chromatography-tandem mass spectrometry (LC-MS/MS). A set of 50 single-donor serum samples were assigned target values for 25-hydroxyvitamin D2 [25(OH)D2] and 25-hydroxyvitamin D3 [25(OH)D3] using reference measurement procedures (RMPs). SRM and PT/EQA samples evaluated included SRM 972a (four levels), SRM 2973, six College of American Pathologists (CAP) Accuracy-Based Vitamin D (ABVD) samples, and nine Vitamin D External Quality Assessment Scheme (DEQAS) samples. Results were received from 28 different laboratories using 20 ligand binding assays and 14 LC-MS/MS methods. Using the test assay results for total serum 25(OH)D (i.e., the sum of 25(OH)D2 and 25(OH)D3) determined for the single-donor samples and the RMP target values, the linear regression and 95% prediction intervals (PIs) were calculated. Using a subset of 42 samples that had concentrations of 25(OH)D2 below 30 nmol/L, one or more of the SRM and PT/EQA samples with high concentrations of 25(OH)D2 were deemed non-commutable using 5 of 11 unique ligand binding assays. SRM 972a (level 4), which has high exogenous concentration of 3-epi-25(OH)D3, was deemed non-commutable for 50% of the LC-MS/MS assays.
Citation
Camara, J. E., Wise, S. A., Hoofnagle, A. N., Williams, E. L., Carter, G. D., Jones, J., ... Sempos, C. T. (2021). Assessment of serum total 25-hydroxyvitamin D assay commutability of Standard Reference Materials and College of American Pathologists accuracy-based Vitamin D (ABVD) scheme and Vitamin D external quality assessment scheme (DEQAS) materials: Vitamin D standardization program (VDSP) commutability study 2. Analytical and Bioanalytical Chemistry, 413, 5067–5084. https://doi.org/10.1007/s00216-021-03470-w
Publisher
Springer
Journal
Analytical and Bioanalytical Chemistry
Research Unit
PubMed ID
PubMed Central ID
Type
Article
Language
Description
The final publication is available at Springer via http://dx.doi.org/10.1007/s00216-021-03470-w
Series/Report no.
ISSN
1618-2642
EISSN
1618-2650
ISBN
ISMN
Gov't Doc
Test Link
Sponsors
Embedded videos